Analyzer Sald2300

Particle size distribution can be calculated using the light intensity distribution pattern of scattered light which is generated from sample particles when laser irradiates them.

Laser Diffraction Particle Size Analyzer
Particle size distribution can be calculated using the light intensity distribution pattern of scattered light which is generated from sample particles when laser irradiates them. This is the main particle size analysis method because it has excellent properties such as a wide measurement range, a short measurement time, and the ability to measure both wet and dry samples. Please refer to page 19 for the measurement principle.
Particle size distribution can have a major effect on the characteristics desired for a given application or objective, or on the performance and quality of a nal product. A particle size analyzer with the ability to precisely measure particle size distribution is an essential tool in today’s laboratory. The SALD-2300 is that tool. With a variety of optional units and application software packages, the SALD-2300 can easily address the application requirements in a variety of industries, including pharmaceuticals, cosmetics, foods, drinks, pigments, paints, ceramics and electronic materials.
FUNCTIONS:
1. Wide particle concentration range from 0.1ppm to 20%.
2. Continuous measurement function at minimum 1-second intervals
3. Wide measurement range from 17nm to 2500 μm
ADVANTAGES:
High resolution:  Accurately detects particle size distribution with five peaks
High reliability
High reproducibility:  Improved stability of the optical system
Flexible data analysis:  Equipped with multiple data analysis modes
High efficiency / High reliability:  Automatic refractive index calculation function eliminates the mistake or trouble of selecting refractive indices.
High sensitivity / High concentration
High analysis capability:  Measurement data from multiple facets
Measurement principle
Laser Diffraction Method
Measurement range
SALD-MS23: 17nm(0.017μm) to 2500μm
SALD-BC23: 17nm(0.017μm) to 400μm
SALD-HC23: 30nm(0.030μm) to 280μm
Measurement Unit
SALD2300
Light source
Red Semiconductor laser (Wavelength 680 nm)
Required power supply
115 or 230 VAC as ordered 100 VA
Operating Environment
Temperature: 10 to 30°C,  Humidity: 20 to 80 % (no condensation)
Light detector
Detector elements for UV semiconductor laser
Total 84 elements (78 forward, 1 side, 5 back)
Sampler
SALD-MS23
Dispersing bath
Capacity: 100/200/300 mL
Batch Cell
SALD-BC23
滚动至顶部